Endpoint:
other: Analytical validation method
Type of information:
experimental study
Adequacy of study:
supporting study
Study period:
2012-10-10 to 2012-11-09
Reliability:
1 (reliable without restriction)
Rationale for reliability incl. deficiencies:
other: The study was conducted in accordance with an acceptable scientific method and in compliance with GLP.

Data source

Reference
Reference Type:
study report
Title:
Unnamed
Year:
2013

Materials and methods

Test guideline
Qualifier:
according to guideline
Guideline:
other: Sanco 3029
Deviations:
no
GLP compliance:
yes

Test material

Constituent 1
Chemical structure
Reference substance name:
Cyclohexanethiol
EC Number:
216-378-7
EC Name:
Cyclohexanethiol
Cas Number:
1569-69-3
Molecular formula:
C6H12S
IUPAC Name:
cyclohexanethiol
Constituent 2
Reference substance name:
Cyclohexylmercaptan
IUPAC Name:
Cyclohexylmercaptan
Test material form:
other: liquid
Details on test material:
- Name of test material (as cited in study report): Cyclohexylmercaptan

- Substance type: monoconstituent substance

- Physical state: clear colourless liquid

- Analytical purity: 99.97%

- Purity test date: 2012-05-31

- Lot/batch No.: 2012-05-1765

- Expiration date of the lot/batch: 2014-06-07

- Stability under test conditions: Stable

- Storage condition of test material: At room temperature in the dark under nitrogen

Results and discussion

Results:
The analytical method is succesfully validated

Any other information on results incl. tables

Specificity

Chromatograms of a test substance solution and the blank accuracy sample are shown in Figure 1and Figure 2 (attached), respectively.

 

The chromatogram of the blank accuracy sample showed no peak at the retention time of the test substance. Since no interferences were detected, the specificity requirements were met and the analytical method was found to be specific for the test substance.

Linearity

The calibration line was constructed using all data points. Figure 3 (attached) illustrates the calibration curve and Table 1 shows the statistical parameters.There was a linear relationship between response and test substance concentration in the range of 3.98 – 598 mg/l (in end solution). Since the coefficient of correlation (r) was > 0.99 and the back calculated accuracies of the data points were in the range 85-115% the calibration line was accepted.

 

Table 1       Statistical parameters of the calibration curve

 

 

Slope

1.61´103

Intercept

2.75´102

Weighting factor

1/concentration2

r

0.9998

 

 

Accuracy and repeatability

The results of the accuracy samples are given in Table 2. Since the mean accuracy at each concentration level fell in the criterion 70-110% and the coefficient of variation was ≤ 20% the analytical method was accepted for the analysis of the test substance in water in the target concentration range of 10 - 1000 mg/l.

Table 2       Accuracy samples

Target concentration
[mg/l]

Nominal concentration
[mg/l]

Analysed concentration
[mg/l]

Accuracy

[%]

Mean accuracy
[%]

Coefficient of variation
[%]

 

 

 

 

 

 

10

9.97

9.48

95

98

7.3

 

9.97

9.90

99

 

 

 

9.97

9.29

93

 

 

 

9.97

11.0

110

 

 

 

9.97

9.32

94

 

 

 

 

 

 

 

 

1000

997

987

99

100

0.55

 

997

1002

100

 

 

 

997

992

100

 

 

 

997

996

100

 

 

 

997

992

100

 

 

 

 

 

 

 

 

Limit of quantification

The limit of quantification (LOQ) was assessed as 10 mg/l in water.

Stability of analytical system and end solutions

The results of the stability of the analytical system and end solutions are given in Table 3. Since the coefficient of variation at both concentration levels was ≤ 20% the analytical system and end solutions were stable over at least a 3.65 hour time interval.

 

Table 3       Stability of the analytical system and end solutions

Nominal concentration
[mg/l]

Elapsed time
[hours]

Coefficient of variation
[%]

 

 

 

3.98

4.50

2.5 (n=3)

 

 

 

15.0

4.36

2.0 (n=3)

 

 

 

598

3.65

1.3 (n=3)

 

 

 

Stability of stock solutions

The coefficient of variation on the response factors of the calibration solutions prepared with fresh and stored stock solutions was 0.31%. Since the value was  10% the stock solutions were stable when stored at room temperature for at least 1 day.

Conclusion

The analytical method was validated for the following parameters:

 

Specificity

specific

Linearity

r = 0.9998

Accuracy

98 and 100%

Repeatability

7.3 and 0.55%

Limit of quantification

10 mg/l

Stability analytical system and end solutions

Stable (3.65 hours)

Stability stock solutions

Stable (1 day)

Applicant's summary and conclusion

Conclusions:
The validation of the analytical method was succesfully performed.