Physical & Chemical properties

Flash point

Currently viewing:

Administrative data

Link to relevant study record(s)

Reference
Endpoint:
flash point of flammable liquids
Type of information:
experimental study
Adequacy of study:
key study
Study period:
2016-02-01 to 2016-02-10
Reliability:
1 (reliable without restriction)
Rationale for reliability incl. deficiencies:
guideline study
Qualifier:
according to guideline
Guideline:
EU Method A.9 (Flash-Point)
Version / remarks:
2008
Deviations:
no
Qualifier:
according to guideline
Guideline:
EPA OPPTS 830.6315 (Flammability)
Version / remarks:
1996
Deviations:
no
Qualifier:
according to guideline
Guideline:
ASTM D93 (Standard test methods for flash point by pensky-martens closed cup tester)
Deviations:
no
Qualifier:
according to guideline
Guideline:
other: ASTM D6450 - 12e1 Standard Test Method for Flash Point by Continuously Closed Cup (CCCFP) Tester
Deviations:
no
Qualifier:
according to guideline
Guideline:
other: ASTM D7094 – 12e1 Standard Test Method for Flash Point by Modified Continuously Closed Cup (MCCCFP) Tester
Deviations:
no
Qualifier:
according to guideline
Guideline:
other: UN Handbook “Recommendations on the Transport of Dangerous Goods - Manual of Tests and Criteria” Rev. 6, 2015, Method 32.4 “Test methods used for determining flash point and viscosity”
Deviations:
no
GLP compliance:
yes (incl. QA statement)
Type of method:
closed cup
Flash point apparatus:
Pensky-Martens apparatus
Key result
Flash point:
71 °C
Atm. press.:
1 013.25 hPa

For the determination of the initial temperature of flash point measurement a screen test was performed with the test item. The sample was heated from 30 °C to 200 °C using 5.5 °C/min heating rate. Based on this preliminary test the flash point of test item was determined to be: 70.4 °C.

In the course of the study three subsequent tests with the test item were carried out. The measured flash points of test item were:

70.4°C, 70.4 °C and 71.4 °C resulting a mean value: 71 °C at 1013.25 hPa.

The repeatability criterion was met because the maximum difference between replicate determinations was not more than 2°C.

Interpretation of results:
GHS criteria not met
Conclusions:
Flash point of the test item was determined to be 71 °C at 1013.25 hPa.
Executive summary:

The purpose of the study was to determine the flash point of test item using Pensky-Martens Closed Cup method. The flash point of the test item was determined to be 71 °C at 1013.25 hPa.

Description of key information

Flash point of the test item was determined to be 71 °C at 1013.25 °C.

Key value for chemical safety assessment

Flash point at 101 325 Pa:
71 °C

Additional information

The purpose of the study was to determine the flash point of test item using Pensky-Martens Closed Cup method. The flash point of the test item was determined to be 71 °C at 1013.25 hPa.